In the base model, total carbon in any matrix is analyzed by utilizing high temperature catalytic combustion technique. Total nitrogen parameter is an optional addition. Proven chemiluminescence technique allows low detection limits.
Samples are loaded in quartz sample boats, where both solid and liquid samples may fit.
Up to 7 g solid sample loading is possible.
A single shot IC Module is built in the analyzer allowing IC analysis in both solid and liquid samples.
Two models are available: TRL-TOC-S is a single shot model with motorized single sample loading feature, while the NEW MODEL TRL-TOC-S-C Series has got automated sample loading feature for 30 samples, the autosampler being integrated on the main unit.
Performance characteristics are the same for both models.
TRL-TOC-L-C Series comply with most international standards including
ISO8245, ISO10694, ASTM D7573-18, Method9060A, EPA415.1, Standard Method5310B, EN13137
TRL-TOC-S -C Specification | |
Method of analysis | High Temperature Catalytic Oxidation |
Carbon Detection | NDIR |
Nitrogen Analysis | with optional TRL-TN |
Nitrogen Detection | Chemiluminescence |
LOD for Carbon in gr | 1ug |
LOD for Nitrogen gr | 1ug |
LOD for Carbon in concentration | 1ppm |
LOD for Nitrogen concentration | 50ppb |
Analysis range for Carbon | 100% |
Analysis range for Nitrogen | 50000ppm |
Type of Sample | Solid and Liquid |
Analyzed Parameters | TC, IC, Optional TN |
Caculated Parameters | TOC |
Furnace Temperature | Adjustable dual furnace, operating between 450-1100°C |
Sample Loading for TC/TN/TOC Analysis | In sample boat |
Sample Loading for IC Analysis | In 40mL vials |
Sample Loading Automation | Full automation for TC analysis |
Manual Sample Loading | Solid and Liquid samples for IC analysis |
IC Analysis | Single shot s/w operation |
Flow Modes | Dual |
Carrier Gas Flow Control | By Mass Flow Control Valve driven via s/w |
Carrier Gas Conditions | 100-1000mL/min, operator adjustable, MFC controlled |
Operating Conditions | 5 to 45°C |
Power | 1100VA, 230VAC |
Dimensions, Analyzer HxWxD | 65x123x63 cm |
Home . Product . Carbon Nitrogen Analysis . TRL-TOC-S / Turkish
Carbon Nitrogen Analyzers
TRL-CN
Carbon and Nitrogen Analyzers
TRL-CN-C
Carbon and Nitrogen Analyzers
TRL-TOC-S
TOC Analyzer for Solid and Liquid Samples
TRL-TOC-S-C
TOC Analyzer for Solid and Liquid Samples
TRL-TOC-L-C
TOC Analyzer for Liquid Samples
TRL-SC-C
Surface Carbon Analyzer
TRL-ROC-C
Residual Organic Carbon, Elemental Carbon, Inorganic Carbon Analyzer